Spore Testing
A standard procedure should be established for the routine evaluation
of each sterilizer using biological indicators. This should be done
at a minimum of once a week. Your personnel should put forth every
effort toward the standardization of sterilizing techniques within
the guidelines established by the equipment manufacturer. Frequent
third party monitoring of the sterilization process by culture tests,
the maintaining of exact standards for the preparation, packaging
and loading of medical supplies in the sterilizers, plus intelligent
and painstaking attention to details are recognized as essential
to the effective sterilization of medical supplies.
Use of Biological Indicator Strips. When using paper spore strips
inoculated with bacterial spores to determine the efficacy of sterilization
processes, care should be taken to use biological indicators which
meet published guidelines for the sterilant and temperature ranges
being tested. It is also necessary to ensure proper placement of
each strip in the portion of the test pack or load most inaccessible
to the Steam, EtO, chemical, or dry heat sterilant. The recommended
procedure is to select one or more of the largest and most dense
packs which normally constitute a routine load of bulk supplies
and insert the glassine envelope containing the inoculated strips
in the center of each pack. Identify the test pack with an indelible
marking instrument, and place the pack on the edge on the bottom
shelf or the sterilizer chamber, near the front. Add the remainder
of the load in a like manner without indicators, and operate the
sterilizer according to the manufacturer's recommendations. Upon
completion of the sterilizing cycle, remove the glassine containing
the biological indicator from the test pack and deliver them along
with an unexposed biological control, to the laboratory for sterility
testing of the strips, i.e. Coastal BioTech Services, Inc..
Biological Indicator Culturing Procedures. We at Coastal BioTech
Services, Inc. will complete the culturing procedures. All tests
for sterility should be conducted by a microbiologist or clinical
technician observing rigid aseptic techniques throughout the procedure.
The glassine test envelope will be aseptically opened and the spore
strip carefully transferred to a tube of sterile culture media or
equivalent. Resistance values and outgrowth rates may vary depending
on the source of the SCDB used for spore recovery. The spore strips
need to be incubated for seven days at 30-35C for dry heat or gas-EO
sterilizer testing and 55-60C for steam or chemiclave sterilizer
testing.
Biological Indicator Interpretation. We at Coastal BioTech Services,
Inc. will observe to tubes daily during the incubation cycle. The
presence of turbidity and/or a color change to yellow indicates
bacterial growth, presumably due to the spores having survived the
sterilization process. According to USPXXII, a biological indicator
spore strip following exposure to gas-EO, chemiclave, or steam at
121C is to have a specific survival and kill times which correspond
to its D value and spore concentration. The survival time is when
100% of the exposed biological indicators demonstrate growth upon
culturing in SCDB and the kill time results in no growth of all
biological indicators exposed.
Biological Indicator Controls. At least one control should be included
with each test series. This requires the transfer of an unexposed
spore strip to a tube of sterile culture media, followed by incubation
at the same temperature corresponding to the test strips. Microbiological
growth indicates that the medium possesses suitable growth promoting
properties and the spore strips contained viable spores prior to
the sterilizing process. Conversely, a microbiological medium control
consisting of one or two tubes of sterile culture media is also
included in the incubation cycle. The absence of growth following
incubation would indicate that the media was sterile prior to the
sterility test procedures.
|